Voyager STR-DE MALDI-TOF-MS
Applied Biosystems |
|

Specifications:
- Sensitivity - Routine detection of 5 fmol of neurotensin with a signal to noise ratio > 80:1.
- Mass Accuracy [Linear Mode] - better than 0.05% for angiotensin [1,296.6853] and myoglobin [16,952.5] with external calibration.
- Mass Accuracy [Reflector Mode] - better than 0.008% for angiotensin and ACTH 18-39 with external calibration.
- Mass Resolution [Linear Mode] - >3,500 for ACTH 18.39 [m/z 2,465.1989], >1,000 for myoglobin [m/z 16,952] and >100 for BSA [m/z 66,431].
- Mass Resolution [Reflector Mode] - >10,000 for insulin (m/z 5,734) and >20,000 for ACTH 18-39.
Instrument Features:
- A high performance Matrix-Assisted Laser Desorption Ionization Time-of-Flight (MALDI-TOF) system which is a proven technology for the MS analysis of both organic and inorganic compounds and biomolecule characterization.
- With its 3-meter effective flight path, reflector mode and patented Delayed Extraction® (DE) technology, the Voyager is still regarded as the industry standard for superior spectral quality, delivering high resolution, sensitivity, and exceptional mass accuracy.
- Voyager Version 5 with Data Explorer™ software permits flexible acquisition and data processing in both positive and negative ion modes of operation to characterize complex mixtures and determine molecular weight on sub-picomole samples.
- Post-source decay (PSD) and collision-induced dissociation (CID) options provide more structural information.
- Fully automated acquisition and processing of data is possible for higher throughput.
|
|